Maximum Range
Voltage
±210V
Current
±1.05 A
Power
22W
Voltage Resolution
1µV
Current Resolution
10pA
Source
DC Voltage
Output Voltage
±21V/ ±1.05 A, ±210V/±105 mA
Current Limit
Min. 0.1% of range
Programming Resolution
1µV, ±200.000mV range
10µV, ±2.00000V range
100µV, ±20.0000V range
1mV, ±200.000V range
Programming Accuracy 1
±(0.02%+600µV), ±200.000mV range
±(0.02%+600µV), ±2.00000V range
±(0.02%+2.4mV), ±20.0000V range
±(0.02%+24mV), ±200.000V range
Load Regulation
0.01% of range + 100µV
Line Regulation
0.01% of range
Overshoot
<0.1% typical (full scale step, resistive load,10mA range)
Recovery Time (1000%Load Change)
<250µs (within 0.1% plus load regulation errors, 1A and 100mA compliance.)
Ripple and Noise
4mV rms(20Hz~ 1MHz) 10mVpp(20Hz~ 1MHz)
Temperature Coefficient
(0°–18°C & 28°–50°C)
±(0.15 × accuracy specification)/°C
DC Current
Output Current
±1.05 A /±21V, ±105 mA /±210V
Voltage Limit
Min. 0.1% of range
Programmed Source Resolution
10pA, ±1.00000µA range
100pA, ±10.0000µA range
1nA, ±100.000µA range
10nA, ±1.00000mA range
100nA, ±10.00000mA range
1µA, ±100.000mA range
10µA, ±1.00000A range
Programmed Source Accuracy 1
±(0.035%+600pA), ±1.00000µA range
±(0.033%+2nA), ±10.0000µA range
±(0.031%+20nA), ±100.000µA range
±(0.034%+200nA), ±1.00000mA range
±(0.045%+2µA), ±10.00000mA range
±(0.066%+20µA), ±100.000mA range
±(0.27%+900µA), ±1.00000A range
Load Regulation
0 0.01% of range + 100pA
Line Regulation
0.01% of range
Overshoot
<0.1% typical (1mA step, RL = 10kΩ, 20V range).
Temperature Coefficient
(0°–18°C & 28°–50°C)
±(0.15 × accuracy specification)/°C
General
Output Settling Time2
100µs typical Time
Output Rise Time (±30%)
300µs, 200V range, 100mA compliance.
150V/µs, 20V range, 100mA compliance.
DC Floating Voltage
Output can be floated up to ±250VDC
Remote Sense
Up to 1V dro per load lead.
Compliance Accuracy
Add 0.3% of range and ±0.02% of reading to base specification.
Range Change Overshoot 3
Adjacent range changes between 200mV, 2V and 20V ranges, 100mV typical.
Minimum Compliance Value
0.1% of range
Command Processing Time 4
Autorange On:10ms.Autorange Off:7ms.
Measurement
Voltage
Input Resistance
>10 G
Measurement Resolution
1µV, ±200.000mV range
10µV, ±2.00000V range
100µV, ±20.0000V range
1mV, ±200.000V range
Measurement Accuracy
±(0.012%+300µV), ±200.000mV range
±(0.012%+300µV), ±2.00000V range
±(0.015%+1.5mV), ±20.0000V range
±(0.015%+10mV), ±200.000V range
Temperature Coefficient
(0°–18°C & 28°–50°C)
±(0.15 × accuracy specification)/°C
Current
Voltage Burden (4-wire mode)
< 1mV
Programmed Source Resolution
10pA, ±1.00000µA range
100pA, ±10.0000µA range
1nA, ±100.000µA range
10nA, ±1.00000mA range
100nA, ±10.00000mA range
1µA, ±100.000mA range
10µA, ±1.00000A range
Programmed Source Accuracy1
±(0.029%+300pA)), ±1.00000µA range
±(0.027%+700pA), ±10.0000µA range
±(0.025%+6nA), ±100.000uA range
±(0.027%+60nA), ±1.00000mA range
±(0.035%+600nA), ±10.00000mA range
±(0.055%+6µA), ±100.000mA range
±(0.22%+570µA), ±1.00000A range
Temperature Coefficient
(0°–18°C & 28°–50°C)
±(0.1 × accuracy specification) / °C
Resistance
Range
Resolution
Test current
Accuracy
<2.00000Ω
---
---
Source IACC+Meas.VACC
2.00000Ω
10µΩ
---
Source IACC+Meas.VACC
20.0000Ω
100µΩ
100mA
±(0.1%+0.003 Ω), Normal
±(0.07%+0.001 Ω), Enhanced
200.000Ω
1mΩ
10mA
±(0.08%+0.03 Ω), Normal
±(0.05%+0.01 Ω), Enhanced
2.00000kΩ
10mΩ
1mA
±(0.07%+0.3 Ω), Normal
±(0.05%+0.1 Ω), Enhanced
20.0000kΩ
100mΩ
100µA
±(0.06%+3 Ω), Normal
±(0.04%+1 Ω), Enhanced
200.000kΩ
1Ω
10µA
±(0.07%+30 Ω), Normal
±(0.05%+10 Ω), Enhanced
2.00000MΩ
10Ω
5µA
±(0.11%+300 Ω), Normal
±(0.05%+100 Ω), Enhanced
20.0000MΩ
100Ω
0.5µA
±(0.11%+1k Ω), Normal
±(0.05%+500 Ω), Enhanced
200.000MΩ
1kΩ
100nA
±(0.66%+10k Ω), Normal
±(0.35%+5k Ω), Enhanced
>200.000M Ω
---
---
Source IACC+Meas.VACC
Temperature Coefficient
(0°–18°C & 28°–50°C)
±(0.15 × accuracy specification)/°C
Source I mode, Manual OHMS
Total uncertainty = I source accuracy + V measure accuracy (4-wire remote sense).
Source V mode, Manual OHMS
Total uncertainty = V source accuracy + I measure accuracy (4-wire remote sense).
6-wire OHMS mode
Available using active ohms guard and guard sense. Max. Guard Output Current: 50mA (except 1A range). Accuracy is load dependent.
Guard Output Impedance
<0.1 in ohms mode
System Speed
Maximum Range Change Rate
75/second
Maximum Measure Auto Range Time
40ms (fixed source) 6
Sequence reading rates7 (rdg./second) for 60Hz (50Hz)
Speed
NPLC/ Trig Origin
Measure
Source-Measure 9
TO MEM.
TO GPIB
TO MEM.
TO GPIB
Fast
0.01 / internal
2081 (2030)
1198 (1210)
1551 (1515)
1000 (900)
488.2
0.01 / external
1239 (1200)
1079 (1050)
1018 (990)
916 (835)
Medium
0.1 / internal
510 (433)
509 (433)
470 (405)
470 (410)
488.2
0.1 / external
438 (380)
438 (380)
409 (360)
409 (365)
Normal
1 / internal
59 (49)
59 (49)
58 (48)
58 (48)
488.2
1 / external
57 (48)
57 (48)
57 (48)
57 (47)
Speed
NPLC/ Trig Origin
Source-Measure Pass/Fail test 8,9
Measure Memory 9
TO MEM.
TO GPIB
TO MEM.
TO GPIB
Fast
0.01 / internal
902 (900)
809 (840)
165 (162)
164 (162)
488.2
0.01 / external
830 (830)
756 (780)
163 (160)
162 (160)
Medium
0.1 / internal
389 (343)
388 (343)
133 (126)
132 (126)
488.2
0.1 / external
374 (333)
374 (333)
131 (125)
131 (125)
Normal
1 / internal
56 (47)
56 (47)
44 (38)
44 (38)
488.2
1 / external
56 (47)
56 (47)
44 (38)
44 (38)
Single Reading Operation Rates (rdg./second) for 60Hz (50Hz)
Speed
NPLC/ Trig Origin
Measure
Source-Measure 9
Source-Measure Pass/Fail test 8,9
TO GPIB
TO GPIB
TO GPIB
Fast(488.2)
0.01 / internal
256 (256)
79 (83)
79 (83)
Medium(488.2)
0.1 / internal
167 (166)
72 (70)
69 (70)
Normal(488.2)
1 / internal
49 (42)
34 (31)
35 (30)
Component Interface Handler Time for 60Hz (50Hz):8, 10
Speed
NPLC/ Trig Origin
Measure
Source Pass/Fail test
Source-Measure Pass/Fail test 9,11
TO GPIB
TO GPIB
TO GPIB
Fast
0.01 / internal
1.04 ms (1.08 ms)
0.5 ms (0.5 ms)
4.82 ms (5.3 ms)
Medium
0.1 / internal
2.55 ms (2.9 ms)
0.5 ms (0.5 ms)
6.27 ms (7.1 ms)
Normal
1 / internal
17.53 ms (20.9 ms)
0.5 ms (0.5 ms)
21.31 ms (25.0 ms)
System General
Load Impedance
Stable into 20,000pF typical
Differential mode Voltage
250 VPk
Common mode Voltage
250 VDC
Common mode Isolation
>10G, <1000pF
Over Range
105% of range, source and measure.
Max. Voltage drop
5V
Max. Sense lead Resistance
1M
Sense input Impedance
>100G
Guard offset Voltage
<150µV, typical
Source output modes
Fixed DC level, Memory List (mixed function), Stair (linear and log)
Source memory list
100 points max.
Memory buffer
5,000 readings @ 5 digits (two 2,500 point buffers). Includes selected measured value(s) and time stamp. Lithium battery backup(3 yr+ battery life).
Programmability
IEEE-488.2 (SCPI)
5 user-definable power-up states plus factory default and *RST.
Digital I/O Connector
Active low input.
Start of test, end of test, 3 category bits.
+5V@ 300mA supply.
1 trigger input, 4 TTL/Relay Drive outputs (33V @ 500mA, diode)
Remote Interface
USB/GPIB/LAN/RS-232
Insulation
Chassis and Terminal
20M or above (DC 500V)
Chassis and AC cord
30M or above (DC 500V)
Operation Environment
Indoor use, Altitude: ≤ 2000m Ambient temperature: 0 ~ 40°C Relative humidity: ≤ 80%
Installation category: II, Pollution degree: 2
Storage Environment
Temperature: -20C ~ 70C
Humidity: < 80%
Input Power
100-240VAC, 50-60Hz
Power Consumption
80W
Accessories
CD User manual x1, Quick Start manual x1
Test lead GTL-207A x 1, GTL-108A x 1, ALLIGATOR CLIP x 2
Dimensions
214 (W) x 86 (H) x 356.5 (D) mm
Weight
Approx. 4.8kg
Remarks
Speed = Normal (1 PLC). For 0.1 PLC, add 0.005% of range to offset specifications, except 200mV, 1A ranges, add 0.05%. For 0.01 PLC, add 0.05% of range to offset specifications, except 200mV, 1A ranges, add 0.5%.
Required to reach 0.1% of final value after Command is processed. Resistive load. 10µA to 100mA range.
Overshoot into a fully resistive 100k load, 10Hz to 1MHz BW, adjacent ranges: 100mV typical, except 20V/200V.
Maximum time required for the output to begin to change following the receipt of :SOURce:VOLTage|CURRent <nrf> Command
Reading rates applicable for voltage or current measurements, autorange off, filter off, display off, trigger delay = 0, and binary reading forma.
Purely resistive lead. 1µA and 10µA ranges <65ms.
1000 point sweep was characterized with the source on a fixed rang.
Pass/Fail test performed using one high limit and one low math limit.
Includes time to re-program source to a new level before making measurement.
Time from falling edge of START OF TEST signal to falling edge of END OF TEST signal.
Command processing time of :SOURce:VOLTage|CURRent: TRIGgered<nrf> Command not included.